DC
Voltage Range & Accuracy |
Range |
DM3559 |
DM3557 |
500.00mV,
5.0000V, |
0.02%
+ 2d |
0.03%
+ 2d |
50.000V |
|
|
500.00V |
0.04%
+ 2d |
0.05%
+ 2d |
Input
Impedance: 10MΩ // 30pF nominal |
(80pF
nominal for 500mV Range) |
Ohms
Range & Accuracy |
Range |
DM3559 |
DM3557 |
500.00Ω |
0.07%
+ 10d |
|
5.0000KΩ |
|
0.1%
+ 6d |
50.000KΩ |
0.07%
+ 2d |
|
5.0000MΩ |
0.2%
+ 6d |
0.4%
+ 6d |
50.000MΩ |
2.0%
+ 6d |
2.0%
+ 6d |
O.C.Voltage
: <1.3VDC (<3VDC for 500Ω) |
Capacitance
Range & Accuracy |
50.00Nf |
0.8%
+ 3d |
500.0nf |
0.8%
+ 3d |
5.000uF |
1.0%
+ 3d |
50.00uF |
2.0%
+ 3d |
500.0uF |
3.5%
+ 5d |
9999uF |
5.0%
+ 5d |
Accuracy
with Film Capacitor or Better |
DC
Current Range & Accuracy |
500.00uA |
0.15%
+ 20d |
5000.0uA |
0.1%
+ 20d |
50.000mA |
0.15%
+ 15d |
500.00mA |
0.1%
+ 20d |
5.0000A |
0.5%
+ 10d |
10.000A
|
*
0.5% + 20d |
*10A
Continuous, 20A for 30 Second Max with |
5
Minutes Cool Down Interval |
Audible
Continuity Tester |
Audible
Threshold: Between 20Ω and 200Ω |
Response
Time < 100uS |
dBm
|
At
600Ω, -11.76dBm to 54.25dBm |
Accuracy:
0.25dB + 2d (@40Hz to 20KHz) |
Selectable
Reference Impedance Values of 4, 8, |
16,
32, 50, 75, 93, 110, 125, 135, 150, 200, |
250,
300, 500, 600, 800, 900, 1000, 1200Ω |
Input
Impedance: 10MΩ // 30pF nominal |
|
AC
/ AC+DC Voltage Range & Accuracy |
|
DM3559 |
DM3557 |
20Hz
to 45Hz |
500.00mV,
5.0000V |
1.5%
+ 40d |
|
50.000V |
|
Unspec'd |
500.00V,
1000.0V |
Unspec'd |
|
45Hz
to 300Hz |
500.00mV |
0.3%
+ 20d |
|
5.0000V,
50.000V |
0.8%
+ 20d |
0.8%
+ 60d |
500.00V,
1000.0V |
0.4%
+ 40d |
|
|
300Hz
to 5KHz |
300Hz
to 1KHz |
500.00mV |
0.3%
+ 10d |
0.8%
+ 40d |
5.0000V,
50.000V, |
0.4%
+ 40d |
2.0%
+ 60d |
500.00V |
|
|
1000.0V |
0.8%
+ 40d |
1.0%
+ 40d |
|
5KHz
to 20KHz |
1KHz
to 20KHz |
500.00mV |
0.5%
+ 20d |
1Db
** |
5.0000V,
50.000V |
0.8%
+ 20d |
2dB
** |
500.00V |
0.5%
+ 20d |
3dB
** |
1000.0V |
Unspec'd |
Unspec'd |
20KHz
to 100KHz |
500.00mV |
2.0%
+ 40d |
|
5.0000V,
50.000V |
4.0%
+ 40d ** |
Unspec'd |
500.00V |
Unspec'd |
|
1000.0V |
|
|
Input
Impedance: 10MΩ // 30pF nominal |
(80pF
nominal for 500mV Range) |
Residual
Reading less than 50d with Test Leads Shorted |
*
From 5% to 10% of Range : |
Accuracy
% of Reading (or in dB) + 80d |
**
From 5% to 10% of Range : |
Accuracy
% of Reading (or in dB) + 180d |
*
*From 10% to 15% of Range : |
Accuracy
% of Reading (or in dB) + 100d |
1000V
Frequency Range to 1KHz |
T1-T2
Dual Temperature (DM3559 Only)
|
Range |
Accuracy |
-50.0℃
to 1000.0℃ |
0.3%
+ 1℃ |
-58.0℉
to 1832.0℉ |
0.3%
+ 2℉ |
Thermocouple
Range & Accuracy not Included |
Diode
Tester |
Range |
Accuracy |
Test
Current |
Open
Circuit |
|
|
Typical |
Voltage |
5.0000V |
1%
+ 1d |
0.8mA |
<
3.5VDC |
DC
Loop Current %4-20mA |
4mA=0%(Zero),
20mA=100%(Span) |
Resolution:
0.01% Accuracy: ±25d |
|
AC
/ AC+DC Current Range & Accuracy |
Range |
DM3559 |
DM3557 |
50Hz
to 60Hz |
500.00uA |
|
|
5000.0uA |
|
|
50.000mA |
0.5%
+ 50d |
1.0%
+ 40d |
500.00mA |
|
|
5.0000A |
|
|
10.000A
* |
|
|
40Hz
to 1KHz |
500.00uA |
|
|
5000.0uA |
|
|
50.000mA |
0.7%
+ 50d |
1.0%
+ 40d |
500.00mA |
|
|
5.0000A |
|
|
10.000A
* |
|
|
1KHz
to 10KHz |
500.00uA |
|
|
5000.0uA |
2.0%
+ 50d |
Unspec'd |
50.000mA |
|
|
500.00mA |
|
|
5.0000A |
Unspec'd |
Unspec'd |
10.000A
* |
|
|
*10A
Continuous, 20A for 30 Second Max with |
5
Minutes Cool Down Interval |
AC
Hz Line Level Frequency |
Range |
Accuracy |
5.0000Hz
to 200.000KHz |
0.002%
+ 4d |
Sensitivities
(Sine Wave): |
0.1V min
|
@500mV
Range |
1V min
|
@5V
Range |
10V min
|
@50V
Range |
100V min
|
@500V
Range |
900V min
|
@1000V
Range |
Logic
Hz Logic Level Frequency |
Range |
Accuracy |
5.0000Hz |
0.002%
+ 4d |
to
2.00000MHz |
|
Sensitivity
: 2.5V Square Wave |
%
Duty Cycle |
Range |
Accuracy |
0.1%
to 99.99% |
3d/KHz
+ 2d |
Input
Frequency: 5Hz to 500KHz, 5V TTL Family |
Crest
Mode (Instantaneous Peak Hold) |
Accuracy:
Specified Accuracy ±100d for changes |
>0.8msec
in Duration |
|