DC Voltage
Range & Accuracy |
| Range |
DM3559 |
DM3557 |
500.00mV,
5.0000V, |
0.02% + 2d |
0.03% + 2d |
| 50.000V |
|
|
| 500.00V |
0.04% + 2d |
0.05% + 2d |
|
Input
Impedance: 10MΩ // 30pF nominal |
|
(80pF
nominal for 500mV Range) |
|
Ohms Range
& Accuracy |
| Range |
DM3559 |
DM3557 |
| 500.00Ω |
0.07% + 10d |
|
| 5.0000KΩ |
|
0.1% + 6d |
| 50.000KΩ |
0.07% + 2d |
|
| 5.0000MΩ |
0.2% + 6d |
0.4% + 6d |
| 50.000MΩ |
2.0% + 6d |
2.0% + 6d |
|
O.C.Voltage
: <1.3VDC (<3VDC for 500Ω) |
|
Capacitance
Range & Accuracy |
|
50.00Nf |
0.8% + 3d |
|
500.0nf |
0.8% + 3d |
|
5.000uF |
1.0% + 3d |
|
50.00uF |
2.0% + 3d |
|
500.0uF |
3.5% + 5d |
|
9999uF |
5.0% + 5d |
|
Accuracy
with Film Capacitor or Better |
|
DC Current
Range & Accuracy |
|
500.00uA |
0.15%
+ 20d |
|
5000.0uA |
0.1% + 20d |
|
50.000mA |
0.15% + 15d |
|
500.00mA |
0.1% + 20d |
|
5.0000A |
0.5% + 10d |
|
10.000A |
* 0.5% +
20d |
|
*10A
Continuous, 20A for 30 Second Max with |
|
5 Minutes
Cool Down Interval |
|
Audible
Continuity Tester |
|
Audible
Threshold: Between 20Ω and 200Ω |
|
Response
Time < 100uS |
|
dBm |
|
At 600Ω,
-11.76dBm to 54.25dBm |
|
Accuracy:
0.25dB + 2d (@40Hz to 20KHz) |
|
Selectable
Reference Impedance Values of 4, 8, |
|
16, 32, 50,
75, 93, 110, 125, 135, 150, 200, |
|
250, 300,
500, 600, 800, 900, 1000, 1200Ω |
|
Input
Impedance: 10MΩ // 30pF nominal |
|
|
AC / AC+DC
Voltage Range & Accuracy |
| |
DM3559 |
DM3557 |
|
20Hz to
45Hz |
500.00mV,
5.0000V |
1.5%
+ 40d |
|
|
50.000V |
|
Unspec'd |
|
500.00V,
1000.0V |
Unspec'd |
|
|
45Hz to
300Hz |
|
500.00mV |
0.3% + 20d |
|
|
5.0000V,
50.000V |
0.8% + 20d |
0.8% + 60d |
|
500.00V,
1000.0V |
0.4% + 40d |
|
| |
300Hz to
5KHz |
300Hz to
1KHz |
|
500.00mV |
0.3% + 10d |
0.8% + 40d |
|
5.0000V,
50.000V, |
0.4% + 40d |
2.0% + 60d |
|
500.00V |
|
|
|
1000.0V |
0.8% + 40d |
1.0% + 40d |
| |
5KHz to
20KHz |
1KHz to
20KHz |
|
500.00mV |
0.5% + 20d |
1Db ** |
|
5.0000V,
50.000V |
0.8% + 20d |
2dB ** |
|
500.00V |
0.5% + 20d |
3dB ** |
|
1000.0V |
Unspec'd |
Unspec'd |
|
20KHz to
100KHz |
|
500.00mV |
2.0% + 40d |
|
|
5.0000V,
50.000V |
4.0% + 40d
** |
Unspec'd |
|
500.00V |
Unspec'd |
|
|
1000.0V |
|
|
|
Input
Impedance: 10MΩ // 30pF nominal |
|
(80pF
nominal for 500mV Range) |
|
Residual
Reading less than 50d with Test Leads Shorted |
|
* From 5%
to 10% of Range : |
|
Accuracy %
of Reading (or in dB) + 80d |
|
** From 5%
to 10% of Range : |
|
Accuracy %
of Reading (or in dB) + 180d |
|
* *From 10%
to 15% of Range : |
|
Accuracy %
of Reading (or in dB) + 100d |
|
1000V
Frequency Range to 1KHz |
|
T1-T2 Dual
Temperature (DM3559 Only) |
|
Range |
Accuracy |
|
-50.0℃
to 1000.0℃ |
0.3%
+ 1℃ |
|
-58.0℉
to 1832.0℉ |
0.3%
+ 2℉ |
|
Thermocouple
Range & Accuracy not Included |
|
Diode
Tester |
| Range |
Accuracy |
Test
Current |
Open
Circuit |
| |
|
Typical |
Voltage |
| 5.0000V |
1% + 1d |
0.8mA |
< 3.5VDC |
|
DC Loop
Current %4-20mA |
|
4mA=0%(Zero),
20mA=100%(Span) |
|
Resolution:
0.01% Accuracy: ±25d |
|
|
AC / AC+DC
Current Range & Accuracy |
| Range |
DM3559 |
DM3557 |
|
50Hz to
60Hz |
| 500.00uA |
|
|
| 5000.0uA |
|
|
| 50.000mA |
0.5% + 50d |
1.0% + 40d |
| 500.00mA |
|
|
| 5.0000A |
|
|
| 10.000A * |
|
|
|
40Hz to
1KHz |
| 500.00uA |
|
|
| 5000.0uA |
|
|
| 50.000mA |
0.7% + 50d |
1.0% + 40d |
| 500.00mA |
|
|
| 5.0000A |
|
|
| 10.000A * |
|
|
|
1KHz to
10KHz |
| 500.00uA |
|
|
| 5000.0uA |
2.0% + 50d |
Unspec'd |
| 50.000mA |
|
|
| 500.00mA |
|
|
| 5.0000A |
Unspec'd |
Unspec'd |
| 10.000A * |
|
|
|
*10A
Continuous, 20A for 30 Second Max with |
|
5 Minutes
Cool Down Interval |
|
AC Hz Line
Level Frequency |
|
Range |
Accuracy |
|
5.0000Hz to 200.000KHz |
0.002% + 4d |
|
Sensitivities
(Sine Wave): |
|
0.1V min
|
@500mV Range |
|
1V min
|
@5V Range |
|
10V min
|
@50V Range |
|
100V min
|
@500V Range |
|
900V min
|
@1000V Range |
|
Logic Hz
Logic Level Frequency |
|
Range |
Accuracy |
|
5.0000Hz |
0.002% + 4d |
|
to 2.00000MHz |
|
|
Sensitivity
: 2.5V Square Wave |
|
% Duty
Cycle |
|
Range |
Accuracy |
|
0.1% to 99.99% |
3d/KHz + 2d |
|
Input
Frequency: 5Hz to 500KHz, 5V TTL Family |
|
Crest
Mode (Instantaneous Peak Hold) |
| Accuracy:
Specified Accuracy ±100d for changes |
| >0.8msec
in Duration |
|