Specifications

Measurement modes

Core loss measurement: Frequency (f) or maximum magnetic flux density (Bm) can be varied.

B-H curve measurement: Frequency (f) or maximum magnetic field (Hm) canbe varied.

Permeability measurement: Frequency (f) can be varied.

Measured items

Core loss measurement: Max. magnetic flux density (Bm), residual magnetic

flux density (Br), max. magnetic field (Hm), coercive force (Hc), core

loss (Pc, Pcv, Pcm), phase (q), amplitude permeability (£ga), hysteresis

loss (Ph), eddy-current loss (Pe).

B-H curve measurement: Max. magnetic flux density (Bm), residual magnetic

flux density (Br), max. magnetic field (Hm), coercive force (Hc), core

loss (Pc, Pcv, Pcm), phase (q), amplitude permeability (£ga),

differential permeability (£gdiff), square ratio (Br/Bm), total magnetic flux change (2f m)

Permeability measurement: Max. magnetic flux density (Bm), max. magnetic

field (Hm), core loss (Pc), phase (q), amplitude permeability (£ga),

inductance (L), complex permeability (£g¡¦, £g¡¨).

Operation modes

Single: One measurement step with one operation.

Continuous: Up to 12-step preset schedules (frequencies or amplitudes) can

be measured continuously with one operation.

Set frequency

10 Hz (optionally 1 Hz) to 10 MHz, 2 digits effective

Accuracy: ¡Ó0.1% for the set value.

Measurement accuracy (23¢XC ¡Ó5¢XC )*

H amplitude measurement (dH):

B amplitude measurement (dB):

Phase accuracy (dq):

When an EI core FE122 is measured with the above combination, the following measurements are available.

At 1 MHz, max. 100 mT (millitesla): Windings = 5.

At 3 MHz, max. 50 mT (millitesla): Windings = 3.

Excitation section

Output voltage ¡Ó50 V max., output current ¡Ó1 A max. ( 1.1 kHz) or ¡Ó4 A

max. (10 Hz (optionally 1 Hz) to 1 kHz).

Output impedance 1 ? + 0.1 £gH, output power 25 VA.

Detection section (standard)

Excitation current detection

System: Voltage drop across two ends of non-inductive

resistance.

Detection sensitivity: ¡Ó50 mA to ¡Ó5 A (1-2-5 steps)

Frequency bandwidth: DC to 100 MHz

Magnetic flux detection

System: Detection of voltage (induced voltage) across

two ends of detection coil.

Detection sensitivity: ¡Ó50 mA to ¡Ó50 V (1-2-5 steps)

Frequency bandwidth: DC to 100 MHz

Storage section

Sampling rate: 512 times the set frequency (synchronized with

the built-in oscillator).

Resolution: 16 bits

Display section

System: 9-inch color CRT

Contents: Measuring condition, characteristic values, B-H

curve, characteristic graphs

Printer (Built-in)

Heat-sensitive dot line system. Paper width approx. 112 mm, printing time no

more than approx. 5 sec.

External interface

GP-IB interface

Power supply

100 V AC (90 ~ 132 V) or 200 V AC (180 ¡V 250 V) models.

Approx. 800 W (100 V AC)

Dimensions & weight

Main unit: Approx. 420W x 350H x 520L mm (excluding the pod),

approx. 34 kg (SY-8232 mainframe + SY-301 measurement unit)

Environment

Operating temperatures +5¢XC to +35¢XC (+10¢XC to +30¢XC when the printer is

used)

Performance guaranteed temperatures

+18¢XC to +28¢XC

Provided accessories

SY-921 measurement pod (1 set), power cords (2), connection cables (1 set),

estimation samples (1 set), adapters for calibration (1 set), Thermal paper (2

rolls), fuses (4), instruction manual (1), dust cover (1), accessory case (1).

Measurement time

Approx. 3 sec./step (when the measured frequency is 1 kHz or more)

Oscillating section

Oscillation frequency: 10 Hz to 10 MHz, sine wave

Output voltage/resistance: 0 to ¡Ó2.5 V/50 ?

Built-in excitation control in 3 steps of approximate value, asymptotic and

demagnetization.

VF ¡Ó(2% + 0.2% x ( ¡V 1))**

Vr

VF IF ¡Ó(0.15¢X + 0.07¢X x ( + ¡V 2) + 0.02¢X)**

Vr Ir

IF ¡Ó(2% + 0.2% x ( ¡V 1))**

Ir

* Defined as the basic accuracy. The

measurement accuracy may differ depending

on the D.U.T. and the measuring condition.

For details, please contact to Iwatsu.

** ¡P VF, IF : Amplitude of Full scale

¡P Vr, Ir : Amplitude of Reading

¡@