APF Series-Programmable AC Power Source - Single Phase (0.5~30kVA)

APF-31010 APF-31015 APF-31020 APF-31030 APF-31045 APF-31060 APF-31075

Programmable AC Power Source - Three Phase/Single Phase (10~75kVA)
           
    APF-31010   APF-31020   APF-31045    
    APF-31015   APF-31030   APF-31060    
            APF-31075    
                 

Model

APF-31010

APF-31015

APF-31020

APF-31030

APF-31045

APF-31060

APF-31075

Capacity (kVA)

10

15

20

30

45

60

75

Circuit Type

 IGBT/PWM type

Input

Phase

Three phase

Wave

Sine wave

Voltage

220V/380V

Voltage range

220V/380V±15%

Frequency range

50Hz±3Hz  or  60Hz±3Hz

Power factor

0.85

Output

Phase

Single phase

Wave

Sine wave

Voltage

Low

5V-150VL-N

High

10V-300VL-N

Frequency

47-63Hz, 50Hz, 60Hz

Frequency regulation

0.01%

Max. current

High (A)

41.7

62.5

83.3

125.0

187.5

250.0

312.5

Low (A)

83.3

125.0

166.7

250.0

375.0

500.0

625.0

System

Power regulation

0.5%

Load regulation

0.5%

Total harmonic distortion (THD)

2%  (linear load)

Efficiency

90%

Response time

2ms

Wave factor

3:1

Protection device

Input no-fuse breaker, output no-fuse breaker, electronic circuit instant trip for over/low voltage,
over current, over load, over temperature and short circuit protection and alarm system

Indicator

Display

LCD 320×240

Voltage

Measure range: 0~300V; analytic degree: (<100V) 0.01V,  (100V) 0.1V; accurate degree: 0.5%FS+5COUNT

Current

Phase power5kW  measure range: 050A; analytic degree: (<10A )0.001A, (10A,<50A) 0.01A; accurate degree: 0.5FS+5COUNT

Phase power>5kW  measure range0700A analytic degree:  0.01Aaccurate degree: 0.5FS+5COUNT 

Power

Phase power5kW   measure range: 05kW; analytic degree: (<1kW) 0.1W , (1kW,5kW) 0.001kW; accurate degree:0.5FS+5COUNT

Phase power>5kW   measure range075kW; analytic degree: 0.01kW; accurate degree: 0.5FS+5COUNT

Frequency

Measure range45500Hz; analytic degree: 0.1Hz; accurate degree: 0.01%FS+2COUNT

Power factor

Measure range: 01; analytic degree: 0.01; accurate degree: 2

Time

Measure range: 1s99h; analytic degree:1s

Default item

Voltage

Fixed range: 0300Vanalytic degree: 0.1Vaccurate degree: 1%

Frequency

Fixed range: 45500Hz analytic degree: 0.1Hzaccurate degree: 0.01

Time

Fixed range: 1s99h; analytic degree:1s

General program memory

10 sets, for voltage and frequency default value

Step program memory

24 sets, for voltage, frequency and step running time

Gradual program memory

12 sets, for voltage, frequency and gradual running time

Revolving operation function

Both step and gradual have revolving function, revolving time can be set 999999 times

Control mode

Short range control

Keyboard control

Remote control

RS232standard

RS485standard

Environment

Insulation resistance

DC500V 10MΩ

Withstand voltage insulation

AC 1800V 10mA/ 1Min

Cooling system

Fan cooling

Temperature

0 ~ 45

Humidity

0~90%  (Non-condensing)

Altitude

1500m

Case No.

Weight(kg)

230

280

305

359

540

620

670

Remarks Custom-made specifications are discussible;

All specifications are subject to change without prior notice.

Products list | brand | Model No. | Home