Instrument Inputs

Input Specifications Channel 1 & 2 (53131A, 53132A)1 Channel 1 (53181A)

Frequency Range
dc Coupled dc to 225 MHz
ac Coupled 1 MHz to 225 MHz (50 Ω)
¡@ 30 Hz to 225 MHz (1 MΩ)
FM Tolerance  25%
¡@ ¡@
Voltage Range and Sensitivity (Sinusoid)2
dc to 100 MHz 20 mVrms to ¡Ó5 V ac + dc
100 MHz to 200 MHz 30 mVrms to ¡Ó5 V ac + dc
200 MHz to 225 MHz 40 mVrms to ¡Ó5 V ac + dc (all specified at 75 mVrms with opt. rear connectors)3
¡@  
Voltage Range and Sensitivity (Single-Shot Pulse)2
4.5 ns to 10 ns Pulse Width 100 mVpp to 10 Vpp (150 mVpp with optional rear connectors)3
>10 ns Pulse Width 50 mVpp to 10 Vpp (100 mVpp with optional rear connectors)3
¡@ ¡@
Trigger Level2
Range ¡Ó 5.125 V
Accuracy ¡Ó (15 mV + 1% of trigger level)
Resolution 5 mV
¡@ ¡@
Damage Level
50 Ω 5 Vrms
0 to 3.5 kHz, 1 MΩ 350 Vdc + ac pk
3.5 kHz to 100 kHz, 1 MΩ 350 Vdc + ac pk linearly derated to 5 Vrms
>100 kHz, 1 MΩ 5 Vrms
¡@ ¡@
Input Characteristics Channel 1 & 2 (53131A, 53132A)1 Channel 1 (53181A)
Impedance 1MΩ or 50Ω
1 MΩ Capacitance 30 pF
Coupling ac or dc
Low-Pass Filter 100 kHz, switchable -20 dB at > 1 MHz
Input Sensitivity Selectable between Low, Medium, or High (default). Low is approximately 2 x High Sensitivity.
¡@
1. Specifications and Characteristics for Channels 1 and 2 are identical for both common and separate configurations.
2. Values shown are for X1 attenuator setting. Multiply all values by 10 (nominal) when using the X10 attenuator setting.

3. When the 53131A or 53132A are ordered with the optional rear terminals (Opt. 060), the channel 1 and 2 inputs are active

on both front and rear of the counter. When the 53181A is ordered with the optional rear terminal, the channel 1 input is
active on both front and rear of the counter. For this condition, specifications indicated for the rear connections also
apply to the front connections.

¡@ ¡@
Trigger Slope Positive or Negative
¡@ ¡@
Auto Trigger Level
Range 0 to 100% in 10% steps
Frequency > 100 Hz
Input Amplitude > 100 mVpp (No amplitude modulation)
¡@ ¡@
Attenuator
Voltage Range x10
Trigger Range x10
¡@ ¡@
Input Specifications4 Channel 3 (53131A, 53132A) Channel 2 (53181A) Frequency Range
Option 015 100 MHz to 1.5 GHz

(for 53181A only)

 (see Opt. 030 for additional specs)
Option 030 100 MHz to 3 GHz
Option 050 200 MHz to 5 GHz
Option 124 200 MHz to 12.4 GHz
¡@ ¡@
Power Range and Sensitivity (Sinusoid)
Option 030 100 MHz to 2.7 GHz: -27 dBm to +19 dBm
¡@ 2.7 GHz to 3 GHz: -21 dBm to +13 dBm
Option 050 200 MHz to 5 GHz: -23 dBm to +13 dBm
Option 124 200 MHz to 12.4 GHz -23 dBm to +13 dBm
¡@ ¡@

4. When optional additional channels are ordered with Opt. 060, refer to configuration table for Opt. 060 under ordering

info on page 8. There is no degradation in specifications for this input, as applicable.

¡@ ¡@
Damage Level
Option 030  5 Vrms
Option 050 +25 dBm
Option 124 +25 dBm
¡@ ¡@
Characteristics
Impedance  50 Ω
Coupling AC
VSWR <2.5:1
¡@ ¡@
External Arm Input Specifications5 Signal Input Range      TTL Compatible
Timing Restrictions
Pulse Width > 50 ns
Transition Time < 250 ns
Start-to-Stop Time > 50 ns
¡@ ¡@
Damage Level 10 Vrms
¡@ ¡@
External Arm Input Characteristics5
Impedance 1 kΩ
Input Capacitance 17 pF
Start/Stop Slope Positive or Negative
¡@ ¡@
External Time Base Input Specifications
Voltage Range 200 mVrms to 10 Vrms
Damage Level 10 Vrms
Frequency 1 MHz, 5 MHz, and 10 MHz (53132A 10 MHz only)
¡@ ¡@
Time Base Output Specifications
Output Frequency 10 MHz
Voltage > 1 Vpp into 50 Ω (centered around 0 V)
¡@ ¡@
Measurement Specifications Frequency (53131A, 53132A, 53181A)
Channel 1 and 2 (53131, 53132); Channel 1 (53181)
Range 0.1 Hz to 225 MHz
¡@ ¡@
Channel 3 (53131A, 53132A) Channel 2 (53181A)
Option 015 (53181only)  100 MHz to 1.5 GHz
Option 030 100 MHz to 3 GHz
Option 050 200 MHz to 5 GHz
Option 124 200 MHz to 12.4 GHz
(Period 2 or 3 selectable via GPIB only)
¡@ ¡@
Period (53131, 53132, 53181) Channel 1 and 2 (53131, 53132); Channel 1 (53181)
Range 4.44 ns to 10 s
¡@ ¡@
Channel 3 (53131A, 53132A); Channel 2 (53181A)
Option 015 (53181 only)  0.66 ns to 10 ns
Option 030 0.33 ns to 10 ns
Option 050 0.2 ns to 5 ns
Option 124 80 ps to 5 ns
¡@ ¡@
Frequency Ratio (53131, 53132, 53181)
Measurement is specified over the full signal range of each input.
Results Range 10-10 to 1011
"Auto" Gate Time 100 ms
¡@ ¡@
5. Available for all measurements except Peak Volts. External Arm is referred to as External Gate for some measurements.

Up | Next | Products list | brand | Model No. | Home