Contact
| Item |
Description |
| GL900-8 |
GL900-4 |
| No. of analog input ch. |
8 ch |
4 ch |
| External input/output |
Trigger input (1 channel), Logic input (4 channels) or Pulse input (4 channels), Alarm output (4 channels) |
| Sampling interval |
10μs to 1 min |
| TIME/DIV |
10 ms/DIV to 24 hour/DIV |
| Timer functions |
Date and time, daily cycle, hourly cycle |
| Trigger functions |
Type |
Start: Data capture starts when a trigger is activated; Stop: Data capture stops when a trigger is activated |
| Condition |
Start: Off, Input signal level (analog, logic/pulse), External*1, Scheduled time |
| Stop: Off, Input signal level (analog, logic/pulse), External *1, Scheduled time |
| Combination |
Input signal level: Level OR, Level AND, Edge OR, Edge AND |
| Mode |
H (Rising), L (Falling), Window In*2, Window Out*2 |
| Alarm setting functions |
Rising, Falling, Window In*2, Window Out*2 |
| Alarm output *1 |
Number of channels: 4, Open collector output (5V, 10 kΩ pull-up resistance) |
| Pulse input*1, *3 |
RPM mode |
50 to 20 M RPM/F.S. (in steps of 1, 2, or 5) |
| Count mode |
50 to 20 M C/F.S. (in steps of 1, 2, or 5) |
| Inst. Mode |
50 to 20 M C/F.S. (in steps of 1, 2, or 5) |
| Calculation functions |
Statistical calculations *4: Average, Peak, Maximum, Minimum, RMS (2 calculations can be set simultaneously) |
| Other functions |
Search function, annotation input function |
| PC inteface |
Ethernet (10BASE-T/100BASE-TX), USB (High Speed supported) provided as standard |
| Ethernet functions |
Web server function, FTP server function, NTP client function |
| USB function |
USB drive mode (File transfer and deletion from internal GL900 memory) |
| Memory device |
Internal |
One million data points / Internal flash memory:Approx. 256 MB |
| External |
USB memory slot (High speed supported) *5 |
| Display screens |
Waveforms + digital values, enlarged waveforms, digital values + calculation results, X-Y |
| Display unit |
5.7-inch TFT color LCD |
| Operating environment |
0 to 40℃, 5 to 85% R.H. (15 to 35℃ when using batteries) |
| Withstand voltage |
Between each input channel and GND: 1000 V p-p for one minute, between input terminals: 1000 Vp-p for one minute |
| Power supply |
AC adapter |
100 to 240 VAC, 50 to 60 Hz |
| DC input |
8.5 to 24 VDC |
| Battery pack *6 |
Option |
| Power consumption |
42VA (when operating and charging battery with AC power) |
40VA (when operating and charging battery with AC power) |
| External dimensions |
232 x 150 x 80 mm (W x H x D), approx. |
| Weight (approx.) |
1.1 kg (excluding AC adapter and battery) |
1.0 kg (excluding AC adapter and battery) |
| Vibration-tested conditions |
Equivalent to automobile parts Type 1 Category A classification |
Terminal block specifications
| Item |
Description |
| GL900-8 |
GL900-4 |
| Number of input channels |
Fixed to 8 channels |
Fixed to 4 channels |
| Input terminal type |
Voltage |
BNC connector |
| Temperature |
M3 screw type terminal board *7 |
| Input method |
All channels isolated Imbalanced input Simultaneous sampling of all channels |
| Measurement ranges |
Voltage |
20, 50, 100, 200, 500 mV; 1, 2, 5, 10, 20, 50, 100, 200, 500 V F.S., 1-5 V F.S. |
| Temperature |
Thermocouples : K, J, E, T, R, S, B, N, W (WRe5-26) |
| Humidity |
0 to 100% (voltage 0 V to 1 V scaling conversion) * with B-530 (option) |
| Input filter |
Off, Line, 5 Hz, 50 Hz, 500 Hz |
Measurement accuracy *8
(23°C±5°C)
When 30 minutes or more have elapsed after power was switched on Filter : Line
GND : connected |
Voltage |
±0.25% of F.S. |
| Thermocouple |
Type |
Measurement temperature range |
Measurement accuracy |
| R/S |
0°C ≦ TS ≦ 100°C |
±7.0°C |
| 100°C < TS ≦ 300°C |
±5.0°C |
| R:300°C < TS ≦ 1600°C |
±(0.05% of rdg +3.0°C) |
| S:300°C < TS ≦ 1760°C |
±(0.05% of rdg +3.0°C) |
| B |
400°C ≦ TS ≦ 600°C |
±5.5°C |
| 600°C < TS ≦ 1820°C |
±(0.05% of rdg +3.0°C) |
| K |
-200°C ≦ TS ≦ -100°C |
±(0.05% of rdg +3.0°C) |
| -100°C < TS ≦ 1370°C |
±(0.05% of rdg +2.0°C) |
| E |
-200°C ≦ TS ≦ -100°C |
±(0.05% of rdg +3.0°C) |
| -100°C < TS ≦ 800°C |
±(0.05% of rdg +2.0°C) |
| T |
-200°C ≦ TS ≦ -100°C |
±(0.1% of rdg +2.5°C) |
| -100°C < TS ≦ 400°C |
±(0.1% of rdg +1.5°C) |
| J |
-200°C ≦ TS ≦ -100°C |
±3.7°C |
| -100°C < TS ≦ 100°C |
±2.7°C |
| 100°C < TS ≦ 1100°C |
±(0.05% of rdg +2.0°C) |
| N |
0°C ≦ TS ≦ 1300°C |
±(0.1% of rdg +2.0°C) |
| W |
0°C ≦ TS ≦ 2000°C |
±(0.1% of rdg +2.5°C) |
Reference contact compensation accuracy : ±1.0°C
‡ Thermocouple diameters T: 0.32mm, others: 0.65mm |
| A/D converter |
16 bits (out of which 14 bits are internally acknowledged) |
| Maximum permissible input voltage |
Between + and - input terminals |
20 mv to 1 V : ±30VDC
2 V to 500 V : ±500VDC |
| Between input channel terminals |
60 Vp-p |
| Between input channel terminal and GND terminal |
60 Vp-p |
| Withstand voltage |
Between input channel terminal and GND terminal |
1 minute at 1000 Vp-p |
| Between input channel terminals |
1 minute at 1000 Vp-p |
-
*1
-
Logic alarm cable (B-513) is required.
-
*2
-
Cannot be set for logic input
-
*3
-
Maximum input frequency: 50 kHz, maximum number of counts: 15 M C
-
*4
-
In real time or when Between Cursors has been specified (during Replay)
-
*5
-
1 file = 2 Gbytes (depends on the USB memory stick used)
-
*6
-
Please install two battery packs.
-
*7
-
Connections are made to both the BNC terminal and M3 screw terminal for the same channel.
-
*8
-
Thermocouple diameters T:0.32φ, others:0.65φ
-
*9
-
Operating temperature range: -25 to +80°C
Control software specifications
| Item |
Description |
| Supported OS |
Windows 2000, Windows XP, Windows Vista (32-bit and 64-bit versions) |
| Functions |
GL900 control, real-time data capture, data conversion |
| Setting range |
Amp settings, data capture settings, trigger settings, alarm settings, other |
| Captured data |
Real-time data |
Binary: Sampling speed: 10 μs to 60 s
CSV: Sampling speed: 10 ms to 60 s |
| Data conversion |
Binary, CSV |
| Display information |
Analog waveforms, logic waveforms, pulse waveforms, digital values |
| File conversion |
Data between cursors, All data |
| 2-screen function (Zoom) |
Display of current and past data |
| Display of statistics and history |
Display of maximum, minimum, and average values |
|
|