Isolation measurement system DM-8000H
The input
block, control block and display block are isolated with optical fiber cables. |
|
Frequency bandwidth: DC to 500MHz. |
|
Simultaneous multi-channel measurement of many channels of different reference potentials. (2 to 24 channels) (DM-900A/AL) |
|
Long-life
battery drive. (The system can be driven by three batteries for about 12
hours) |
|
Measurement using long memory. (DM-900A/AL, DM-910A/AL) |
|
Simultaneous
measurements of the inverter's switching waveform and ON-voltage. |
|
*Distribution
of DM-8000H series are limited in Japanese and Asian market.
Example: Isolation Unit DM-900A x 12units (24ch)
Lineup
|
*1. |
Performance is improved when using three or more units with a memory length exceeding 100k points. |
*2. |
With insulation case. |
*3. |
With insulation case. Optional probe is required for voltage measurements. |
*4. |
Optional cable set without sheath. |
*5. |
Standard item for isolation unit. |
Specification
DM-8000H Main Unit
*When the DM-610A acquisition control card is mounted
Item |
DM-8000H |
Transceiver card connection
Number of slots |
6(Max.12 isolation units and/or acquisition units can be connected.) |
Time axis
Sweep range |
1 ns/div to 20 s/div |
Clock accuracy |
10 ppm |
Acquisition mode |
Normal/peak |
Trigger system
Trigger operation mode |
Auto, Normal, Single, Stop |
Trigger source |
Up to 24 channels |
Trigger type |
Edge or pulse width |
Trigger delay |
Available |
Interface
Ethernet port |
3 port (10BASE-T/100BASE-TX/1000BASE-T) |
Power supply unit
AC power |
100 V to 240 V (50/60 Hz) |
|
Power consumption |
130 VA max |
Mechanical unit
Dimensions and weight |
132 (H) x 351 (W) x 420 (D) mm, Approx. 6.9 kg |
Operating temperature |
0 to +40°C |
Performance guaranteed temperature |
+10 to +35°C |
Accessories
LAN cable |
1 |
Power supply cable |
1 |
Operating manual |
CD-R (1) |
Control software |
IS viewer DM-800 CD-R (1) |
DM-900A,AL/DM-910A,AL Isolation Unit
Item |
DM-900A, DM-900AL |
DM-910A, DM-910AL |
Signal input unit
Frequency Bandwidth(-3dB) |
500 MHz |
||||||
Bandwidth limiter |
20 MHz, 100 MHz |
||||||
Input impedance |
1 Mohm // 16 pF |
||||||
Maximum input voltage |
400 Vmax (DC + AC peak <= 5 kHz) CAT I |
||||||
Number of channels |
2 (between channels are not isolated) |
1 |
|||||
Input coupling |
GND/
DC 1 Mohm, |
GND/ DC 1 Mohm |
|||||
Input sensitivity |
2 mV/div to 10 V/div, 1-2-5 steps |
|
|||||
Offset range |
2
mV/div to 50 mV/div: ±1 V *1 |
|
|||||
DC gain accuracy |
± (1.5% + 0.5% of ful-scale) |
||||||
Offset accuracy |
±
(1.0% + 0.5% of full-scale + X) |
||||||
Probe sense |
10:1, 100:1, 1000:1 (Auto detection/manual settings) |
||||||
Sample rate |
1 GS/s (2 GS/s during interleave) |
||||||
Vertical axis resolution |
8 bits |
||||||
Maximum memory length |
DM-900A:
500k points per channel |
DM-910A:
500k points per channel |
Trigger system unit
Trigger sources |
CH1, CH2 |
CH1-MAIN |
Trigger slope |
Positive / Negative |
|
Coupling |
AC, DC, HFREJ, LFREJ |
|
Level range |
125% of full scale |
Interface
Interface |
One set of three optical interfaces (optical fiber: 2 to 200 m) |
Power supply and battery unit
Internal battery |
3 battery packs (unit can operate on one battery) |
Battery charging |
Can be charged by the main unit |
Battery charger |
Optional |
Power consumption |
120 VAmax (when using AC power) |
Battery operation time |
Approx. 12 hours of continuous operation (when using 3 batteries) |
Battery charging time |
Approx. 6 hours |
AC power supply |
AC 100 to 240 (50/60 Hz) |
Calibration signal
Calibration signal |
0.6 V/6 V (switchable) |
Mechanical unit
Dimensions (mm) |
122.4 (H) x 258.4 (W) x 544 (D) |
Weight |
Appro. 7 kg (excluding battery packs and
accessories) |
Operating temperature |
0 to +40°C |
Performance guaranteed temperature |
+10 to +35°C |
Accessories
Battery pack |
3 |
Power supply cable |
1 |
DM-600A Transceiver Card
Item |
DM-600A |
Number of isolation/acquisition units connected |
2 (DM-900A/AL, DM-910A/AL) |
Operation indicator |
Status display via LED |
Mechanism |
Card inserted in main unit (DM-8000H) |
Operating temperature |
0 to +40°C |
Performance guaranteed temperature |
+10 to +35°C |
IS Viewer DM-800 (supplied with the DM-8000H main unit)
* IS viewer is installed in the controlling computer (option) and is used to operate the isolation system and to monitor waveforms.
Item |
DM-800 |
Main function
Operations |
+, -, ×, ÷, |×|, |÷|, ∫, dy/dx |
Parameter measurements |
Max, Min, p-p, Top, Base, Top-Base, RMS, CycleRMS, Mean, CycleMean, +/-Overshoot, TransitionTime, dv/dt, Freq, Period, +/-PulseCount, +/-PulseWidh, Duty, Integral, Integral(abs), Integral(pos), Integral(neg), Skew(%), Skew(Level) |
Other functions |
XY display, FFT, Cursor, smoothing, channel deskewing, rescaling, offline viewer function |
Saving waveforms |
CSV |
Saving images |
BMP, PNG, Clipboard |
Saving setup values |
With waveforms and without waveforms |
Controlling computer
CPU |
Intel® Pentium®4 Processor or later |
RAM |
2 GB or larger |
OS |
Windows® XP Professional SP3 *Note1 |
Display |
At least WXGA (1280 x 768 pixels)
recommended |
Note 1. |
Intel and Pentium are registered trademarks or trademarks of Intel Corporation and its subsidiary companies in the United States of America and other countries. |
Note 2. |
Windows is a registered trademark or trademark of Microsoft Corporation in the United States of America and other countries. |
Products list | brand | Model No. | Home