DS-5424

DS-5422

DS-5414

DS-5412

Frequency bandwidth(-3dB)

200MHz

100MHz

Rise time(Typical)

750ps

1ns

Input Channel

4

2

4

2

Maximum Sampling Speed

1GS/s(all CHs)/ 2GS/s(Channel combined)

1GS/s(all CHs)

Equivalent Sampling Rate

100GS/s

Peak detect resolution

1ns

Averaging

2 to 256 times

Maximum Memory Length

500kpts/ch(for all CHs)

Vertical Resolution

8-bit

Input voltage range

2mV/div - 10V/div(1MΩ)

Offset voltage

2mV/div - 50mV/div : ±1V
50.2mV/div - 500mV/div : ±10V
502mV/div - 10V/div : ±100V

DC gain accuracy

±1.5% + (0.5% of fullscale)

Maximum Input Voltage

±400Vpeak CAT I (1MΩ)

Band-limiting filter

Analog form : 20MHz,2MHz,200kHz

Input coupling

GND, DC 1MΩ, AC 1MΩ

Input Impedance

1MΩ±1.5% //20pF

Probe sense

Automatic: 11101100110001
Manual
: 11511012015011001200150011000120001

Timebase

2ns/div 50s/div

5ns/div 50s/div

Standard probe

SS-0130R (multi-channel)

Roll mode

50ms/div 50s/div (100kS/s max)

Timebase (Clock) accuracy

±10 ppm

Trigger function

Trigger type

Edge, Pulse Count, Pulse Width, Period, Dropout, TV

TV mode /
Line number /
Field sequence

NTSC, PAL, Custom / upto 3,000 / 1, 2, 4 & 8 fields

Pulse Count Trigger

1 to 9999 events

Pulse Width Trigger

15ns to 50s

Period Trigger

40ns to 50s

Dropout Trigger

50ns to 50s

Pattern Trigger

-

Trigger source /
Status /
threshhold level setting

-

Trigger Source

Input CHs, Line, EXT(+/-0.5V), EXT10(+/-5.0V)

Trigger Slope /Coupling

+,- / AC, DC, HF Rej, LF Rej, Noise Rej

Display function

Display size/ resolution

7.5inch color TFT-LCD with Touch screen / VGA(640*480pixels)

Display mode

YTXYXY (Triggered)

Vector Display Method

Interpolation on sample points or Dots

Analog persistence mode

Monochrome grayscale or Color Spectrum

Persistence time setting

100ms, 200ms, 500ms, 1s, 2s, 5s, 10s & Infinite

Reference waveform memory

5 waveforms

Panel setting memory

5settings for Internal memory or USB memory

Parameter measurement, Cursor, Zoom, Calculation, Replay

rameter measurement

Maximum, Minimum, Peak-Peak, RMS, Cycle RMS, Mean, Cycle Mean, Top, Base, Top-Base, +Overshoot, -Overshoot, Tr 20-80%, Tf 80- 20%, Tr 10-90%, Tf 90-10%, Freq., Period, +Pulse Count, -Pulse Count, +Pulse Width, -Pulse Width, Duty Cycle, Integral, Skew (+, -), Skew at level

Concurrent number of measurements /
statistics display

Four parameter maximum /Max (Maximum), Min (Minimum) ,Num (Number of the total waveforms)

Logging item, destination

-

Pass/Fail

-

Cursor

Time, Amplitude, Time and Amplitude, Value at cursor

Zoom

Zoom key enable display at individual grid area

Calculation

Addition, Subtraction, Multiplication, FFT(8k points maximum, RECTANGULAR, HANNING, FLATTOP)

Rescaling / Unit conversion

a*x+b(x: input voltage at User defined a and b) / Volt, Ampere, Watt, degree and unit-less

Replay

Automatic waveform recording upto 1,024waveforms, History Replayable

Counter

6-digit

Interface (standard)

USB 2.0(Host&Device), LAN(100Base-TX), GPIB(Factory option: DS-576)

AUX Interface
(Options)

 

AUX connector for External options

DS-577 AUX IO CH1/CH2 output

-

DS-578 AUX IO CH1/TRIG output

-

DS-576 GPIB interface

GPIB: IEEE488.2 (factory option)

DS-579 probe power supply options

Our active probe power supply (2 systems)

Waveform Data Storage

USB memory for Binary, ASCII, Mathcad, Calculation(ASCII) & Calculation(Mathcad)

Hardcopy Output

Output form front panel USB port to USB memory in TIFF, BMP & PNG format or Output to PictBridge Printers

Calibration signal output

Square waveform at 1kHz, 3Vp-p

Power / power consumption

AC 90Vrms to 240Vrms, 47Hz to 63Hz, AC 90Vrms to 132Vrms, 380Hz to 420 Hz, 95VA max (60W max)

Dimensions (mm) / Weight

Approx. 330W × 190H ×124D / Approx. 3.7kg

Environmental condition

Performance guarantee temperature

+10℃ to + 35℃

Operation temperature and humidity /
Altitude conditions

0deg. to +40˚C at 5 to 80%(RH<=30˚C) and 55%RH or less at 40˚C non-condensation /
2,000meter or lower

Accessory

Probe(multi-channel), Power cord, Front panel cover, CD-ROM(Instruction Manual, Remote Control Manual), User's Guide

Factory options

GPIB Interface
DS-576
(factory option)

GPIB: IEEE488.2 (factory option)

2 ways optional for Probe Power supply
DS-579

Our active probe power supply (2 systems)

 

 