Newtons4th N4L Impedance Analysis Interface: Impedance Analyzer
 
 

Contact

Impedance Analysis Interface: Impedance Analyzer

PSM series accessory for LCR test

The PSM1700 and PSM1735 can provide LCR measurements in either 'passive mode', where an external shunt is used for current detection or with optional impedance measurement accessories available from N4L.

Passive mode

In this mode, an external shunt is placed in series with the component under test, then CH1 is used to measure the voltage across the component under test while CH2 is used to measure the voltage across the shunt. From these voltages and their relative phase angle, the PSM units compute impedance and all LCR functions.

LCR Active Head and Impedance Analysis Interface

The LCR Active Head and IAI convert the PSM1700/1735 units into high performance LCR meters with true 4 wire Kelvin connections that are taken directly to the component under test without the need for external shunts. Buffering, amplification and selectable shunts provide LCR measurements over a wide frequency and impedance range.

Impedance Analysis Interface
connected to the PSM1735

IAI Kelvin Fixture

 

  LCR Active Head IAI
Frequency Range 10£gHz to 5MHz 10£gHz to 35MHz
Measurement Ranges    
Inductance 100nH to 10kH
Capacitance 10pF to 1000£gF
Resistance 10m£[ to 100M£[
Absolute accuracy 0.2% <1kHz 0.1% <1kHz
  0.5% + 0.005%/kHz <5MHz 0.2% + 0.0005%/kHz <35MHz

Accuracy with reference

component trim

0.2% + 0.0002%/kHz  
Internal Shunts:    
Low N/A 5£[
Normal 100£[ 50£[
High 10k£[ 5k£[
Very High 1M£[ 500k£[
Power source Auxiliary Port 90¡V264V rms 47¡V64Hz

LCR measurement display

Downloads

Brochure & Technical Specifications

PDF, 130Kb

This product is designed, manufactured & supported in the UK. All measurements are traceable to UKAS, certificates of conformance and calibration are supplied as standard. Our products are CE marked.